Processor Preventative Maintenance Program
GRAND X-RAY ensures the quality of our preventative maintenance program by maintaining a
database of information about your piece of equipment. This will include the installation
date, serial number, model number as well as time cycle, replenishment rates, developer
and water and dryer temperatures.
The following is a summary of GRAND X-RAY's monthly cleaning services:
|
|
Detector, crossover and squeegee
assemblies are removed for cleaning and inspection |
|
|
Deep racks are removed for
cleaning and inspection |
|
|
Solutions are drained, tanks are
cleaned and inspected and system cleaner is used as needed |
|
|
Developer filter is cleaned or
changed |
|
|
Tanks are filled with fresh
solutions, circulation of fixer and developer is checked and roller transport racks,
crossover, etc., are checked and replaced as necessary |
|
|
Developer, water and dryer
temperatures are checked and compared to service card and are adjusted as needed |
|
|
Replenishment rates are checked,
compared to the service record and adjusted as needed |
|
|
Test films are run to check the
full operation of the machine to ensure there are no marks or scratches |
|
|
Clean all processor surfaces,
outer tank walls, inner side panels, base plate and covers |
|
|
Check operation of all dryer
rollers and tubes and clean for chemical deposits and dust. Any worn gears, pulleys,
bearings, o-rings or belts will be replaced as necessary |
|
|
|
|
|
|




 |
|